Appendix B: Troubleshooting
Device Front Panel ACCESS LED on PXI Module is Off
When PXI Chassis is On
If the ACCESS LED is not lit after you power on the PXI chassis, a problem may exist with the
PXI power rails, a hardware device, or the LED.
Note
The LEDs may not light until the device has been configured in MAX. Before
troubleshooting this issue, verify that the device appears in MAX.
Troubleshoot this issue by completing the following steps:
1.
Power off your PXI chassis.
2.
Disconnect any signals from the PXI module front panel.
3.
Remove the PXI module and inspect for signs of damage. Do
not
reinstall a damaged
device.
4.
Reinstall the PXI module using the procedure described in step
5. Installing the Hardware
.
5.
Power on your PXI chassis.
6.
Verify the device appears in MAX.
7.
Reset the device in MAX and perform a self-test. Refer to step
6. Configuring and Testing
in MAX
for information about performing device resets and self-tests in MAX.
8.
If the ACCESS LED still fails to light, contact NI support at
ni.com/support
.
Device Does Not Appear in MAX
Complete the following steps if the NI device does not appear in MAX:
1.
In the MAX Configuration pane, click
Devices and Interfaces
to expand the category.
2.
Click
the appropriate Chassis
and press <F5> to refresh the list of installed devices.
3.
If the device is still not listed, power off the system, ensure the device is correctly installed,
and restart.
4.
If the device still does not appear under
Devices and Interfaces
, contact NI support at
ni.com/support
.
Device Failed the Self-Test
The MAX self-test performs a brief test of device resources. If the device does not pass the
self-test, complete the following steps:
1.
Reboot your system.
2.
Launch MAX and perform the self-test again. If the device still fails the self-test, proceed
to step 3.
3.
Uninstall and reinstall NI-HSDIO.
4.
If the device still fails the self-test, contact NI support at
ni.com/support
.