Chapter 3
Configuring and Testing
3-4
ni.com
Shunt Calibration Switches
Shunt calibration switches A and B are software configurable settings that
allow you to engage or disengage the shunt calibration resistors in order to
perform gain calibration. In most cases, you do not explicitly control the
shunt calibration switches, but instead allow driver software to
automatically adjust them for you during the automated shunt calibration
procedure. However, if you want to explicitly control the calibration
switches, you can write an application program that controls the shunt
calibration switches. Refer to Chapter 4,
information.
Note
Null calibration is done for you automatically if you perform shunt calibration using
the NI-DAQmx driver. Refer to Chapter 4,
, for more information
about how to perform shunt calibration using the driver.
Simultaneous Sample and Hold
When it is critical to measure two or more signals at the same instant
in time, simultaneous sample and hold (SS/H) is required. Typical
applications that might require SS/H include vibration measurements and
phase difference measurements. You can disable this setting through your
application if you require scan rates beyond the maximum allowable with
SS/H engaged. NI recommends leaving SS/H engaged. Disabling SS/H
introduces a small offset voltage. You can compensate for this offset by
performing offset null calibration. Refer to Chapter 5,
, for more information about how to enable and disable SS/H.
Note
You cannot change the simultaneous sampling mode in MAX. You must use an ADE
such as LabVIEW to configure the setting. Refer to your ADE help file for more
information.
Configurable Settings in MAX
Note
If you are not using an NI ADE, using an NI ADE prior to version 7.0, or are using
an unlicensed copy of an NI ADE, additional dialog boxes from the NI License Manager
appear allowing you to create a task or global channel in unlicensed mode. These messages
continue to appear until you install version 7.0 or later of an NI ADE.
You can use MAX to configure your bridge-based sensor measurement.
This section describes where you can access each software-configurable
setting available in MAX.