Appendix A
Specifications
©
National Instruments Corporation
A-5
Shock and Vibration
Operational shock .................................. 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random Vibration
Operating................................................ 5 to 500 Hz, 0.3 g
rms
Nonoperating.......................................... 5 to 500 Hz, 2.4 g
rms
(Tested in accordance with
IEC-60068-2-64. Nonoperating
test profile exceeds the
requirements of
MIL-PRF-28800F, Class 3.)
Safety
The PXI-1050 is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
•
EN 61010-1, IEC 61010-1
•
UL 61010-1
•
CAN/CSA-C22.2 No. 61010.1
Note
For UL and other safety certifications, refer to the product label, or visit
ni.com/certification
, search by model number or product line, and click the
appropriate link in the Certification column.
Electromagnetic Compatibility
Emissions ............................................... EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
1
Immunity ................................................ EN 61326:1997 + A2:2001,
Table 1
EMC/EMI............................................... CE, C-Tick, and FCC Part 15
(Class A) Compliant
1
DC input exempt from emissions requirements.