CALIBRATION PROCEDURE
NI 9265
4-Channel, 0–20 mA, 16-Bit Analog Current Output Module
This document contains the verification and adjustment procedures for the National Instruments
9265. For more information on calibration solutions, visit
ni.com/calibration
.
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 3
Verification............................................................................................................................... 4
Accuracy Verification....................................................................................................... 4
Adjustment................................................................................................................................ 5
Accuracy Adjustment ....................................................................................................... 5
EEPROM Update ..................................................................................................................... 6
Reverification ........................................................................................................................... 6
Accuracy Under Calibration Conditions .................................................................................. 6
Worldwide Support and Services ............................................................................................. 7
Software
Calibrating the NI 9265 requires the installation of NI-DAQmx 9.6 or later on the calibration
system. You can download NI-DAQmx from
ni.com/downloads
. NI-DAQmx supports
LabVIEW, LabWindows
™
/CVI
™
, ANSI C, and .NET. When you install NI-DAQmx, you only
need to install support for the application software that you intend to use.