The following accuracy table is valid for calibration under the following conditions:

Ambient temperature 23 °C ± 5 °C

NI 9251 with mini XLR installed in slot 8 of an NI cDAQ-9178 chassis

Slots 1 through 7 of the NI cDAQ-9178 chassis are empty

Table 6. NI 9251 with mini XLR Accuracy under Calibration Conditions

Device

Percent of Reading (Gain Error)

Percent of Range

1

 

 

 

NI 9251 with mini XLR

±0.1198%

±0.0680%

Note

 

For operational specifications, refer to the most recent

NI 9251 with mini XLR Datasheet

 online at 

ni.com/manuals

.

Worldwide Support and Services

The NI website is your complete resource for technical support. At 

ni.com/support

, you have

access to everything from troubleshooting and application development self-help resources to
email and phone assistance from NI Application Engineers.

Visit 

ni.com/services

 for NI Factory Installation Services, repairs, extended warranty, and

other services.

Visit 

ni.com/register

 to register your NI product. Product registration facilitates technical

support and ensures that you receive important information updates from NI.

A Declaration of Conformity (DoC) is our claim of compliance with the Council of the
European Communities using the manufacturer’s declaration of conformity. This system
affords the user protection for electromagnetic compatibility (EMC) and product safety. You
can obtain the DoC for your product by visiting 

ni.com/certification

. If your product supports

calibration, you can obtain the calibration certificate for your product at 

ni.com/calibration

.

NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas,
78759-3504. NI also has offices located around the world. For telephone support in the United
States, create your service request at 

ni.com/support

 or dial 1 866 ASK MYNI (275 6964). For

telephone support outside the United States, visit the 

Worldwide Offices

 section of 

ni.com/

niglobal

 to access the branch office websites, which provide up-to-date contact information,

support phone numbers, email addresses, and current events.

1

Range equals 4.243 V.

NI 9251 Calibration Procedure 

  | 

© National Instruments

  | 

7

Summary of Contents for NI 9251

Page 1: ...e 6 Reverification 6 Accuracy under Calibration Conditions 6 Worldwide Support and Services 7 Software Calibrating the NI 9251 with mini XLR requires the installation of NI DAQmx 16 1 or later on the calibration system You can download NI DAQmx from ni com downloads NI DAQmx supports LabVIEW LabWindows CVI ANSI C and NET When you install NI DAQmx you only need to install support for the applicatio...

Page 2: ...nd NI DAQmx C properties NI DAQmx NET Help Support for Visual Studio Reference information for NI DAQmx NET methods and NI DAQmx NET properties key concepts and a C enum to NET enum mapping table Test Equipment The following table lists the equipment recommended for the performance verification and adjustment procedures If the recommended equipment is not available select a substitute using the re...

Page 3: ...copper wire for all cable connections to the device Use twisted pairs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The device temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the device measurement circuitry is at a stable operating temperature Initial Setup...

Page 4: ...mine the As Found status of the NI 9251 with mini XLR 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9251 with mini XLR to the calibrator as shown in the following figure Figure 1 Analog Input Verification Connections NI 9251 Calibrator AI0 AI0 Chassis GND HI LO AI1 AI1 Note The chassis ground lug connection to ground is the NI 9251 with mini XLR ground connection 3 Set the calibrator ...

Page 5: ...Disconnect the calibrator from the device Table 4 NI 9251 with mini XLR Verification Test Limits Test Point 2 Year Limits Location Value V Lower Limit VDC Upper Limit VDC Minimum 1 9 1 905161 1 894839 Mid scale 0 0 002885 0 002885 Maximum 1 9 1 894839 1 905161 Note The test limits listed in the NI 9251 with mini XLR Verification Test Limits table are derived using values from the NI 9251 with mini...

Page 6: ...point in the array i Close the calibration session 4 Disconnect the NI 9251 with mini XLR from the calibrator EEPROM Update When an adjustment procedure is completed the NI 9251 with mini XLR internal calibration memory EEPROM is immediately updated If you do not want to perform an adjustment you can update the calibration date and onboard calibration temperature without making any adjustments by ...

Page 7: ...t Product registration facilitates technical support and ensures that you receive important information updates from NI A Declaration of Conformity DoC is our claim of compliance with the Council of the European Communities using the manufacturer s declaration of conformity This system affords the user protection for electromagnetic compatibility EMC and product safety You can obtain the DoC for y...

Page 8: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relevant HTS codes ECCNs and other import export data NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY...

Reviews: