Figure 2. NI 9246 Pinout
AI0+
AI1+
AI2+
AI1-
AI0-
AI2-
Note
The NI 9246 does not measure DC currents.
Caution
Connecting a DC current source to the NI 9246 outside of the
specifications stated in the NI 9246 Datasheet may lead to invalid
measurements. Refer to the datasheet on
for the NI 9246
specifications.
Connecting Cables Using Ring Lug Connectors
Complete the following steps to connect cables using ring lug connectors. Use 5.26
mm
2
(10 AWG) cables.
© National Instruments
7
NI-9246 Getting Started