NI 6738/6739 Calibration Procedure
|
© National Instruments
|
25
5.
Use the DMM to read the value of AO 0.
6.
Call
DAQmxClearTask
with the following parameter:
•
taskHandle
:
taskHandle
7.
Call
DAQmxInitExtCal
with the following parameters:
•
deviceName
:
Dev_name
•
password
:
NI
•
calHandle
:
&calHandle
8.
Call
DAQmx
AO
SeriesCalAdjust
with the following parameters:
•
calHandle
:
calHandle
•
referenceVoltage
: Value read from DMM
9.
Call
DAQmxCloseExtCal
with the following parameters:
•
calHandle
:
calHandle
•
action
:
DAQmx_Val_Action_Commit
This function also saves the date, time, and temperature of the adjustment to the onboard
memory.
Note
If an error occurs during adjustment, no constants are written to the
EEPROM.
10. Disconnect the DMM from the device.
EEPROM Update
When an adjustment procedure is completed, the NI 6738/6739 internal calibration memory
(EEPROM) is immediately updated.
If you do not want to perform an adjustment, you can update the calibration date without making
any adjustments by initializing an external calibration (using the
DAQmxInitExtCal
function
call) and closing the external calibration (using the
DAQmxCloseExtCal
function call).
Reverification
Repeat the
section to determine the as-left status of the device.
Note
If any test fails reverification after performing an adjustment, verify that you
successfully self-calibrated the device, as described in the
before returning your device to NI. Refer to
Worldwide Support and Services
assistance in returning the device to NI.