background image

NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

2

ni.com

External Calibration

External calibration requires a high-precision digital multimeter (DMM) 
to measure error on the NI 6703/6704 outputs. The error is then corrected 
with a set of calibration constants that adjust the offset and gain of the 
16-bit DAC to ensure that the output voltages and currents are within the 
device specifications.

Because of the NI 6703/6704 calibration circuitry, the only calibration 
adjustment you must make is to the values of the calibration channels to 
account for time- or temperature-related drift of the onboard reference.

Equipment and Other Test Requirements

This section describes the equipment, test conditions, documentation, and 
software you need to calibrate the NI 6703/6704.

Test Equipment

To calibrate the NI 6703/6704, you need a high-precision DMM that is at 
least 10 ppm (0.001%) accurate. NI recommends that you use the 
Hewlett-Packard 3458A DMM for calibration.

If you do not have a Hewlett-Packard 3458A DMM, use the accuracy 
specifications to select a substitute calibration standard.

If you do not have custom connection hardware, you may need a connector 
block such as the NI CB-68 and a cable such as the SH6868-D1. These 
components give you easy access to the individual pins on the 68-pin 
I/O connector.

Test Conditions

Follow these guidelines to optimize connections and test conditions during 
calibration:

Keep connections to the NI 6703/6704 short. Long cables and wires 
act as antennae, picking up extra noise that can affect measurements.

Use shielded copper wire for all cable connections to the device. 
Use twisted-pair wire to eliminate noise and thermal offsets.

Maintain a temperature between 18 and 28 °C. To operate the module 
at a specific temperature outside this range, calibrate the device at that 
temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 15 minutes to ensure that the 
measurement circuitry is at a stable operating temperature.

Summary of Contents for NI 6703

Page 1: ...6704 products have two calibration options an internal or self calibration and an external calibration Internal Calibration The NI 6703 6704 is continuously applying corrections to the output levels b...

Page 2: ...e NI recommends that you use the Hewlett Packard 3458A DMM for calibration If you do not have a Hewlett Packard 3458A DMM use the accuracy specifications to select a substitute calibration standard If...

Page 3: ...constants update the calibration date and write to the factory calibration area The factory calibration area and the calibration date should only be modified by a metrology laboratory or another faci...

Page 4: ...ode If you do not want to use these variable definitions you can examine the function call listings in the NI DAQ documentation and the nidaqcns h file to determine what input values are required Docu...

Page 5: ...e channel 0 is operating within specifications all the voltage channels should be operating within specifications If you want to verify the channels individually this procedure can be easily expanded...

Page 6: ...from the NI 6703 6704 You have now verified the voltage output of the NI 6703 6704 Verifying Current Output NI 6704 Only This procedure verifies that the current output of channel 16 is within specif...

Page 7: ...rom the NI 6704 You have now verified the current output of the NI 6704 Adjusting the NI 6703 6704 At the end of each calibration procedure new constants are stored in the user calibration area of the...

Page 8: ...age channel 0 Decreasing the voltage value increases the magnitude of the voltage on voltage channel 0 6 Call AO_VWrite to generate 10 0 V on voltage channel 0 Set the following parameters chan 0 volt...

Page 9: ...Call AO_VWrite to generate 0 1 mA on current channel 16 Set the following parameters chan 16 voltage 0 1 5 Call AO_VWrite with the channel set to the current offset channel Adjust the voltage paramet...

Page 10: ...values should also be stored in the factory calibration area Calibration constants stored in the factory calibration area cannot be overwritten by the user To save the calibration constants and updat...

Page 11: ...mn contains the upper limits and lower limits for the test point value If the NI 6703 6704 has been calibrated in the last 24 hours the test point value should be between the upper and lower limit val...

Page 12: ...ructure Voltage Output Verification Figure 1 Voltage Output Verification Table 2 Current Output Specifications NI 6704 Only Test Point Value mA 24 Hour Ranges 1 Year Ranges Lower Limit mA Upper Limit...

Page 13: ...6704 Only Figure 2 Current Output Verification NI 6704 Only Refer to Table 2 Current Output Specifications NI 6704 Only to determine the appropriate ranges for the voltages and currents to verify Con...

Page 14: ...voltage channel 0 AO_VWrite deviceNumber 0 0 0 NI 6703 AO_VWrite deviceNumber 16 voltage Adjust voltage until the DMM reads 0 0 V NI 6704 AO_VWrite deviceNumber 32 voltage Adjust voltage until the DMM...

Page 15: ...Output Adjustment NI 6704 Only AO_VWrite deviceNumber 35 voltage Adjust voltage until the DMM reads 20 0 mA Set_DAQ_Device_Info deviceNumber ND_CALIBRATION_ENABLE ND_YES Connect the multimeter to curr...

Page 16: ...ents covering National Instruments products refer to the appropriate location Help Patents in your software the patents txt file on your CD or ni com patents 2002 2004 National Instruments Corp All ri...

Reviews: