1-22
|
ni.com
Chapter 1
Getting Started with the cDAQ Chassis
•
Triggering Modes
—The cDAQ chassis supports different trigger modes, such as start
trigger, reference trigger, and pause trigger with analog, digital, or software sources. Refer
to the following sections for more information:
–
The
Analog Input Triggering Signals
section of Chapter 2,
–
The
Analog Output Triggering Signals
section of Chapter 3,
–
The
Digital Input Triggering Signals
–
The
Digital Output Triggering Signals
section of Chapter 4,
•
Independent Data Streams
—The cDAQ chassis supports seven independent high-speed
data streams, which allow for up to seven simultaneous hardware-timed tasks, such as
analog input, analog output, buffered counter/timers, and hardware-timed digital
input/output.
•
PFI Signals
—The PFI signals provide access to advanced features such as triggering,
synchronization, and counter/timers. You can also enable a programmable debouncing
filter on each PFI signal that, when enabled, samples the input on each rising edge of a filter
clock. PFI signals are available through parallel digital input and output modules installed
in up to two chassis slots and through the two PFI terminals provided on the cDAQ chassis.
Refer to the
section of Chapter 4,
, for more information.
•
Flexible Counter/Timers
—The cDAQ chassis includes four general-purpose 32-bit
counter/timers that can be used to count edges, measure pulse-widths, measure periods and
frequencies, and perform position measurements (encoding). In addition, the
counter/timers can generate pulses, pulse trains, and square waves with adjustable
frequencies. You can access the counter inputs and outputs using parallel digital I/O
modules installed in up to two slots, or by using the two chassis PFI terminals provided on
the cDAQ chassis. Refer to Chapter 5,
, for more information.