1-2
|
ni.com
Chapter 1
Getting Started
Safety Guidelines for Hazardous Voltages
If
hazardous voltages
are connected to the device/module, take the following precautions. A
hazardous voltage is a voltage greater than 42.4 V
pk
or 60 VDC to earth ground.
Caution
Ensure that hazardous voltage wiring is performed only by qualified
personnel adhering to local electrical standards.
Caution
Do
not
mix hazardous voltage circuits and human-accessible circuits on
the same module.
Caution
Make sure that chassis and circuits connected to the module are properly
insulated from human contact.
Caution
NI 62
xx
devices and modules provide no isolation.
Electromagnetic Compatibility Guidelines
This product was tested and complies with the regulatory requirements and limits for
electromagnetic compatibility (EMC) as stated in the product specifications. These requirements
and limits are designed to provide reasonable protection against harmful interference when the
product is operated in its intended operational electromagnetic environment.
This product is intended for use in industrial locations. There is no guarantee that harmful
interference will not occur in a particular installation, when the product is connected to a test
object, or if the product is used in residential areas. To minimize the potential for the product to
cause interference to radio and television reception or to experience unacceptable performance
degradation, install and use this product in strict accordance with the instructions in the product
documentation.
Furthermore, any changes or modifications to the product not expressly approved by National
Instruments could void your authority to operate it under your local regulatory rules.
Caution
To ensure the specified EMC performance, product installation requires
either special considerations or user-installed, add-on devices. Refer to the product
installation instructions for further information.
Caution
For compliance with Electromagnetic Compatibility (EMC)
requirements, this product must be operated with shielded cables and accessories. If
unshielded cables or accessories are used, the EMC specifications are no longer
guaranteed unless all unshielded cables and/or accessories are installed in a shielded
enclosure with properly designed and shielded input/output ports.
Caution
This product may become more sensitive to electromagnetic disturbances
in the operational environment when test leads are attached or when connected to a
test object.