© National Instruments
|
B-11
Figure B-11.
Sample Clock Timebase Timing and the Analog Input Timing Engine
Table B-5.
Convert Clock Timebase Timing
Time
Description
Line
Min (ns)
Max (ns)
t
13
Delay to Selected Start Trigger
PFI
3.4
8.8
RTSI
3.3
8.5
STAR
2.7
5.7
t
14
Selected Start Trigger Setup Time
(to Sync Convert Clock Timebase)
—
1.5
—
t
15
Selected Start Trigger Hold Time
(to Sync Convert Clock Timebase)
—
0
—
t
16
Sync Convert Clock Timebase to
Start Trigger
—
0.9
2.4
t
17
Start Trigger to POUT
PFI
1.1
3.1
RTSI
1.1
2.7
S
t
a
rt Trigger
Termin
a
l
S
elected Reference Trigger
Reference Trigger
Termin
a
l
Termin
a
l
S
elected
Sa
mple Clock
Termin
a
l
Termin
a
l
Termin
a
l
S
elected
S
t
a
rt Trigger
RT
S
I
Termin
a
l
Termin
a
l
Termin
a
l
S
elected P
aus
e Trigger
S
I
Co
u
nter
Block
S
I2
Co
u
nter
Block
S
I_TC
Sa
mple Clock Time
bas
e
S
ync
Sa
mple Clock Time
bas
e
Convert Clock Time
bas
e
S
ync Convert Clock Time
bas
e
S
I
S
t
a
rt
p_AI_Convert
S
t
a
rt
1
_i
S
I2_TC
P
aus
e Trigger