NI 4065 Calibration Procedure
82
ni.com
29. Call the niDMM Cal Adjust Offset VI.
30. Reset the Calibrator.
Adjusting AC Current
To adjust AC current for the NI 4065, complete the following steps:
1.
Refer to Table 19 for the appropriate calibrator output and parameter
values as you complete the following steps:
a.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance
of this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
b.
Call the niDMM Control VI.
Note
Configure the DMM for current mode before applying a current from the calibrator.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
1
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
:
Cal Session
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_DC_CURRENT
Range
:
1.0
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_Control
with the following
parameters:
Instrument_Handle
:
Cal Session
Control Action
:
Commit