NI 4065 Calibration Procedure
Adjusting AC Voltage
To adjust AC voltage for the NI 4065, complete the following steps:
1.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance of
this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
2.
Call the niDMM Control VI.
3.
Refer to Table 18 for the appropriate calibrator output and parameter
values as you complete the following steps:
a.
On the calibrator, output the value listed in the
Calibrator Output
column in Table 18 for the current iteration.
Allow the calibrator
output to settle before proceeding
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
:
Cal Session
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_AC_VOLTS
Range
:
300
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_Control
with the following
parameters:
Instrument_Handle
:
Cal Session
Control Action
:
Commit