3-4
Radiated Emission and Immunity testing
The calculation of the output level is given by:
where:
m
V
dB
E
μ
=Generated Electric field
(
)
V
dB
G
μ
= Signal Generator Level
( )
dB
C
1
= Interconnecting cable loss
( )
dB
A
= Power amplifier gain
( )
dB
C
2
= Interconnecting cable loss
m
dB
TAF
m
1
3
= Transmit Antenna factor at 3 m =
(
)
5
,
41
1
log
20
3
10
−
−
m
dB
AF
f
m
MHz
The variables and terms in the above expression are used for calibration
test setup only. They demonstrate how the instrumentation and facility
factors contribute to meet the typically required E-field uniformity value of
-0.0 dB to + 6,0 dB.
Fig. 3-2
Radiated immunity test setup
(
)
( ) ( )
( )
+
−
+
−
=
m
dB
TAF
dB
C
dB
A
dB
C
V
dB
G
m
V
dB
E
m
1
3
2
1
μ
μ
Antenna Pow er
am plifier
3 m
Field s trength
m eter
R F s ignal generator
Fiber optic link
C able C able
TAF
C 1
C 2
G
A