
Radiated Emission and Immunity testing
3-1
3 - Radiated Emission and
Immunity testing
3.1 Radiated Emission
testing
The electromagnetic emissions emanating from the equipment under test
(EUT) are measured during a radiated emission test
According to the international Standards, the purpose of this test is to verify
that the EUT's emissions are below specified limits during operation.
The receiving antenna middle point can be located either at 3 or 10 meter
from the EUT and must scan from 1 to 4 meters in height. This scanning
helps the operator to locate the EUT's worst case electromagnetic
emission level.
The typical emissions test set-up is composed of:
•
receive
antenna;
•
interconnecting
cable;
•
preamplifier;
•
preamplifier
cable;
•
meter
receiver.
3.1.1 The Antenna Factor
The performance of the antenna relating to the value of the incident’s field
to the voltage output of the antenna is the Antenna Factor; this value is
provided in dB/m by the manufacturer.
3.1.2 The
interconnecting
cable
The cable connects the antenna to the preamplifier or meter receiver input.
There is a loss in the cable resulting in a reduction of the measured signal,
to increase the measure accuracy, these losses need to be added to the
measured value.
3.1.3 The preamplifier
If the Receiver or the Spectrum Analyzer have a high input noise figure,
may be necessary to compensate it with a preamplifier installed between
the antenna and the meter receiver. The preamplifier makes the measured
signal larger, thus the gain of the preamplifier must be subtracted from the
measured value to obtain the correct final result.
3.1.4 The preamplifier
interconnecting
cable
An additional cable to connect the preamplifier to the meter receiver input
may be necessary if the preamplifier is installed.
There is a loss in the cable resulting in a reduction of the measured signal,
to increase the measure accuracy, these losses need to be added to the
measured value.
Document ASEN-81109 -
NARDA 2018
Summary of Contents for PMM AS02
Page 24: ...1 20 General information Fig 1 17 Typical BL 01 VSWR...
Page 28: ...1 24 General information Fig 1 20 Typical LP 02 VSWR...
Page 32: ...1 28 General information Fig 1 25 Typical LP 04 VSWR...
Page 34: ...1 30 General information Fig 1 27 Typical DR 01 VSWR...
Page 48: ...2 12 Mounting Instruction This page has been left blank intentionally...
Page 54: ...3 6 Radiated Emission and Immunity testing This page has been left blank intentionally...
Page 56: ...Suggerimenti Commenti Note Suggestions Comments Note...