11 Instrument maintenance
120
NBM-550
Narda
11.4 Checking instrument functions
Note:
The function test does not apply to the EHP-50. The
EHP-50 has its own built in self test.
Performing a function test:
1. Connect the probe to the instrument (see
2. Switch the instrument on (see
page 30) and wait until the self test is complete
.
3. Open the
Probe Test
function (MAIN/INFORMATION/
PROBE TEST).
The measured field strength will be displayed separa-
tely for all 3 input channels
(Probe channel 1...3).
4. Bring an appropriate signal source (see
on page 122) near the probe head.
WARNING
Probe is not operating properly
Possibly present high radiation values are not
recognized.
Check probes for proper operation with a signal source
before using this measuring instrument. This is
especially important for thermocouples because the
sensors can be affected by various mechanical and
environmental stressors. Narda offers portable
sources to accomplish this important step
(see
Before beginning any RF radiation measurement,
always advise yourself of the frequencies and field
strengths that you could expect to encounter.
Summary of Contents for NBM-550
Page 10: ...Contents VIII NBM 550 Narda ...
Page 38: ...3 Preparing the NBM 550 for use 28 NBM 550 Narda ...
Page 52: ...4 Getting started 42 NBM 550 Narda ...
Page 88: ...6 Measurements with the NBM 550 and EHP 50 78 NBM 550 Narda ...
Page 110: ...9 PC software 100 NBM 550 Narda ...
Page 139: ...12 10 CE Declaration of Conformity Narda NBM 550 129 12 10 CE Declaration of Conformity ...
Page 140: ...12 Specifications 130 NBM 550 Narda ...
Page 142: ...12 Specifications 132 NBM 550 Narda ...
Page 146: ...13 Ordering information 136 NBM 550 Narda ...
Page 154: ...Index 144 NBM 550 Narda ...
Page 155: ......