17 Specifications
258
IDA
-
3106
Narda
17.3.7 Multi-Channel Power
17.4
Antenna handle and antennas
17.4.1 Antenna handle
Measurement principle
Spectrum analysis, followed by Channel Power evaluation
Number of channels
1 to 500, to be defined on instrument or by IDA Tools PC software
Channel band width CBW,
(-3 dB nominal)
Individually selectable for each channel, from 40 Hz to 6 GHz
Roll-off factor
< 4 x RBW / CBW
Applied RBW
• Automatic: Frequency range of channel divided by 4 but not
more than 20 MHz.
• Manually: Can be set in the range of available RBWs of
spectrum analysis, but not more than CBW of channel
divided by 4.
• Individual: separately defined for each individual service
minimum service bandwidth divided by 4 (Other needs to be off)
Name of channel
• User definable, maximum 15 characters
• Automatic creation of communication channel lists by PC
software. Channel name is assigned automatically.
Detection
Root mean square value (RMS), integration time = 1 / RBW
Trace, RBW
See spectrum analysis mode
Display: Table view
• Channel name, corresponding frequency band, measurement
result, RBW if set individually for each channel.
• Sort function according to columns.
Display: Bar graph view
Bar graph for measurement result of each channel.
Noise threshold
• Selectable at 0, 3, 6, 10, 15, or 20 dB relative to device noise
floor.
• Measurement values below the threshold are shown as
“
<
absolute threshold value”.
Frequency range
• 9 kHz to 6 GHz
• Frequency response correction is applied automatically when
used in conjunction with the IDA basic unit.
Preamplifier
• Built in, can be switched off
• Amplification 20 dB, noise figure < 6 dB
Compass
• Embedded electronic compass,
Azimuth uncertainty < 1.5° RMS (typ.) for tilt < 15°
• Pitch- and roll uncertainty < 3° RMS (typ.) in the range of
30°
(RMS means the standard deviation of the specified error)
Connection cable to
IDA basic unit
RF cable and control cable combined in a flexible tube,
length of 1 meter
Summary of Contents for IDA
Page 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Page 8: ...Contents VI IDA 3106 Narda...
Page 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Page 134: ...6 Common Functions 126 IDA 3106 Narda...
Page 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Page 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Page 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Page 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Page 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Page 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Page 246: ...15 PC Software 238 IDA 3106 Narda...
Page 254: ...16 Remote Control 246 IDA 3106 Narda...
Page 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Page 272: ...17 Specifications 264 IDA 3106 Narda...
Page 284: ...19 Glossary 276 IDA 3106 Narda...
Page 290: ...Index 282 IDA 3106 Narda...
Page 291: ......