10 Time Domain (Scope) Mode
184
IDA
-
3106
Narda
10.3
Selecting measurement parameters
The parameters described in must be set to match the channel being
measured to obtain meaningful results. Fcent and CBW must be set first of
all.
The time span, time resolution, and number of measured values (n) are
linked and can be represented by the following equation:
Time span = Number of measured values (n) x Time resolution
At least 16 measured values must be displayed over the time span. This
relationship is clear from the following table.
Some examples of channels and the appropriate settings of these
parameters are also given in the section
Measurement parameter examples
Table 33: Measurement parameters and their functions
Parameter
Function / Setting
Fcent
Center frequency of channel under observation
Time Span
Time Resolution
Measurement duration for capturing the signal characteristic
Time resolution for the displayed measurement values
CBW
Channel bandwidth of the signal to be displayed
Trace / Detector
The trace types / detectors that can be selected depend on the
measurement function: Act / +Peak, RMS, -Peak
Table 34: Relationship between Time Span and Time Res parameters
Sub mode
Time Span
Time Res
No. of
values (n)
Trace / Detector
High Resolution
Scope
settable,
n / CBW
fixed,
1 / CBW
max. 250,000
Act
Long-Time
Scope
settable,
24 h
settable,
Time Span / n
max. 62,500
+Peak, RMS,
-Peak
Summary of Contents for IDA
Page 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Page 8: ...Contents VI IDA 3106 Narda...
Page 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Page 134: ...6 Common Functions 126 IDA 3106 Narda...
Page 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Page 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Page 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Page 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Page 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Page 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Page 246: ...15 PC Software 238 IDA 3106 Narda...
Page 254: ...16 Remote Control 246 IDA 3106 Narda...
Page 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Page 272: ...17 Specifications 264 IDA 3106 Narda...
Page 284: ...19 Glossary 276 IDA 3106 Narda...
Page 290: ...Index 282 IDA 3106 Narda...
Page 291: ......