7 Direction Finding Mode
138
IDA
-
3106
Narda
7.4.4
Horizontal Scan and manual determination
Direction Finding (Horizontal Scan: On)
Once the bearing is displayed on the polar diagram, you can also specify
the bearing manually. This may be necessary if the calculated result does
not appear to be sensible due to interference.
To set the bearing manually:
You have made a Horizontal Scan.
The red (lower) arrow indicates the calculated bearing. The green
(upper) arrow indicates the direction in which the antenna is pointing.
1.
Point the antenna in the desired direction and press the
Manual Correct. of Direction
softkey.
The red arrow moves to the manually specified direction (green
arrow) when you press the softkey.
2.
Press the
SAVE
key on the Basic Unit to save the measurement results.
A window shows the save process.
Summary of Contents for IDA
Page 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Page 8: ...Contents VI IDA 3106 Narda...
Page 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Page 134: ...6 Common Functions 126 IDA 3106 Narda...
Page 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Page 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Page 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Page 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Page 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Page 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Page 246: ...15 PC Software 238 IDA 3106 Narda...
Page 254: ...16 Remote Control 246 IDA 3106 Narda...
Page 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Page 272: ...17 Specifications 264 IDA 3106 Narda...
Page 284: ...19 Glossary 276 IDA 3106 Narda...
Page 290: ...Index 282 IDA 3106 Narda...
Page 291: ......