6.6 Setting the reference level (Reference Level)
Narda
IDA
-
3106
97
3.
Use the rotary control knob to select the desired setting and press the
OK
key.
The selected value is applied. The selected T-RMS is shown in the
lower status bar.
6.6
Setting the reference level
(Reference Level)
All operating modes
To set the reference level:
Press the
Reference Level
softkey.
You can now select the reference level as follows:
•
Manual reference level selection
•
Automatic reference level search (RL Search)
Background information:
The sensitivity of the system depends on the input attenuator setting.
This setting is determined by the
Reference Level (RL)
parameter. High
measurement sensitivity avoids falsification of the results due to noise
generated internally by the device. On the other hand, it is important that
the device is not overloaded (saturated). Saturation can also be caused
by signals outside the frequency band, e.g. high power radio transmitters
when measuring mobile telephone signals.
The reference level setting covers up to 50 dB depending on the attached
components (handle, type of antenna, etc.) and is coupled to the setting
of the input attenuator (Attenuator, Atten). The input attenuator can be
matched ideally to the measurement signal by means of its fine
adjustment in steps of 1 dB. At the highest reference level setting,
Atten = 50 dB; at the most sensitive reference level, Atten = 0 dB. The
actual value of the input attenuator setting is shown in brackets after the
reference level.
It is possible to decouple the reference level and the input attenuator
indirectly using the
Y-Scale Ref
parameter in the
Display
menu.
Summary of Contents for IDA
Page 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Page 8: ...Contents VI IDA 3106 Narda...
Page 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Page 134: ...6 Common Functions 126 IDA 3106 Narda...
Page 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Page 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Page 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Page 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Page 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Page 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Page 246: ...15 PC Software 238 IDA 3106 Narda...
Page 254: ...16 Remote Control 246 IDA 3106 Narda...
Page 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Page 272: ...17 Specifications 264 IDA 3106 Narda...
Page 284: ...19 Glossary 276 IDA 3106 Narda...
Page 290: ...Index 282 IDA 3106 Narda...
Page 291: ......