10.8 Display of overdriven measured values
Narda
FieldMan
107
10.8 Display of overdriven measured values
Overdriving can lead to damage or even destruction of a measuring probe. In addition,
during overdriving, the measurement results are undervalued due to the occurring com
-
pression and thus recorded as too low. Measurement data stored in data records are
identified by override symbols.
A currently overdriven state is indicated by a red instead of a yellow display field in con
-
nection with an upward pointing triangle symbol
Due to previous overdriving, the statistical measured values of
MAX
,
MAXAVG
and
AVG
are also undervalued and are therefore displayed in red after overdriving has occurred.
In this case, the values in the measurement settings should be reset.