Chapter 15 Advanced Settings
4. Proceed to setting the date, if required.
5. Alternately, press the
Esc
key to return to the
First
Menu Screen
. Press
Esc
again to return to the
measurements screen.
To set the date...
1. Press
F3
(
SET
). The display will change to show six
underscore lines where the new date must be entered.
2. Enter the new date in DD/MM/YY format using the
numeric keypad. If you make a mistake, simply enter
more numbers until the correct time has been entered.
3. Press
F4
(
SAVE
) to save the new time.
4. Proceed to setting the time, if required.
5. Alternately, press the
Esc
key to return to the
First
Menu Screen
. Press
Esc
again to return to the
measurements screen.
Temperature Correction
This feature is
only used when the meter and probe are operating
at significantly different ambient temperatures.
This situation can occur when the meter is located indoors and
the probe is outdoors. Either a probe extension cable or a fiber
optic link system is normally used for such applications.
Temperature correction circuitry is used in the probe and there
is a temperature sensor located inside the meter. To achieve the
highest level of accuracy, it is important that the sensor inside
the meter sense the ambient temperature that the probe is
subjected to. If the two components are operating in different
ambient temperature environments, the 8718B can be set to
T I M E
I S
D A T E
I S
1 3 : 4 2
_ _ / _ _ / _ _
2 4 H R - F O R M A T - D D / M M / Y Y
S A V E
F1
F2
F3
F4
Summary of Contents for 8718B
Page 9: ...8 Operating the Meter...
Page 37: ...36 Operating the Meter...
Page 71: ...70 Operating the Meter...
Page 75: ...74 Operating the Meter...
Page 79: ...78 Operating the Meter...
Page 87: ...86 Operating the Meter...
Page 104: ...103 Chapter 18 User s Software Main Menu Screen...
Page 108: ...107 Chapter 18 User s Software Probe Installation Screen...
Page 112: ...111 Chapter 18 User s Software Logged Survey Data Screen...
Page 124: ...123 Appendix B Theory of Operation Model 8718B Functional Block Diagram...
Page 129: ...128 Appendices...
Page 133: ...132 Appendices...