68
Operating the Meter
4. The left side of the display gives the status of the fixed
alarm. It indicates whether the fixed alarm is either
on
or
off
and the current alarm threshold setting in terms
of percent of full scale of the probe in use (1% to 99%).
5. The right side of the display indicates whether the
variable tone alarm is either
on
or
off
.
6. The bottom line of the display indicates which function
key (
F1
-
F4
) is used to turn the two alarms
on
or
off
.
Fixed Alarm
1. Use
F1
(
ON
) or
F2
(
OFF
) to change the status of the
fixed alarm.
2. Use the numeric key pad to enter a two-digit alarm
threshold number. This number is equal to the percent
of full scale of the measurement range of the probe in
use. For example, to set the meter to sound its alarm
at 5 mW/cm
2
when using a Model 8721D probe, which
has a full scale measurement range of 20 mW/cm
2
, set
the alarm threshold to 25%.
3. If you want to turn the variable tone alarm
on
or
off
do
that next. See
Variable Tone Alarm
on the following
page.
4. When you are satisfied with all the alarm settings, press
ENTER
and the meter will return to the measurement
screen.
N
ote
If you change to an-
other probe with a
different full scale
measurement range,
the alarm threshold
will also change. A
setting that is often
used with shaped fre-
quency response
probes is 50% of
Standard. To set
the alarm for one of
the 8722 series
probes, which have
full scale measure-
ment ranges of 300%
of standard, calculate
the threshold setting
by dividing the de-
sired alarm level by
the full scale value.
In this case
50/300 = 16.66%.
Set the alarm thresh-
old to 17% and the
alarm threshold will
be 51% of Standard
when using a
8722 series probe.
Summary of Contents for 8718B
Page 9: ...8 Operating the Meter...
Page 37: ...36 Operating the Meter...
Page 71: ...70 Operating the Meter...
Page 75: ...74 Operating the Meter...
Page 79: ...78 Operating the Meter...
Page 87: ...86 Operating the Meter...
Page 104: ...103 Chapter 18 User s Software Main Menu Screen...
Page 108: ...107 Chapter 18 User s Software Probe Installation Screen...
Page 112: ...111 Chapter 18 User s Software Logged Survey Data Screen...
Page 124: ...123 Appendix B Theory of Operation Model 8718B Functional Block Diagram...
Page 129: ...128 Appendices...
Page 133: ...132 Appendices...