Chapter 10 Alarm Settings
10
Alarm Settings
Alarm Description
The Model 8718B Meter is equipped with an audible alarm that
emits sound through a tiny hole on the bottom of the meter. The
alarm sounds automatically whenever the field strength
exceeds the measurement range of the probe or
over ranges
.
The meter can also be set to sound its alarm at a specific level
and/or to sound a variable tone indication of field strength.
Controlling the Alarm Functions
To make changes to the audio alarm settings...
1. Access the menu system by pressing the function key
that corresponds to the legend below
MENU
on the
bottom line of the display.
2. Press
8
(
NEXT
) to go to the
Second Menu Screen
.
3. Press
5
(
ALARM
) to access the alarm menu. A new
menu will appear that allows you to:
•
Turn the audio alarm
on
or
off
.
•
Set the alarm threshold in terms of percent of full scale
of the probe in use (1% to 99%).
•
Turn the variable tone alarm
on
or
off
.
N
ote
If you access the
menu system from
the Opening Menu
Screen, you will re-
turn to that screen.
Once an alarm set-
ting has been made,
the alarm function
will be initiated once
the meter is placed
into the measure-
ment mode.
A U D I O
A L A R M
S E T T I N G S
F I X E D
I S
T O N E
I S
O N
@
2 5 %
O F F
O N
O F F
O N
O F F
F 1
F2
F 3
F 4
Summary of Contents for 8718B
Page 9: ...8 Operating the Meter...
Page 37: ...36 Operating the Meter...
Page 71: ...70 Operating the Meter...
Page 75: ...74 Operating the Meter...
Page 79: ...78 Operating the Meter...
Page 87: ...86 Operating the Meter...
Page 104: ...103 Chapter 18 User s Software Main Menu Screen...
Page 108: ...107 Chapter 18 User s Software Probe Installation Screen...
Page 112: ...111 Chapter 18 User s Software Logged Survey Data Screen...
Page 124: ...123 Appendix B Theory of Operation Model 8718B Functional Block Diagram...
Page 129: ...128 Appendices...
Page 133: ...132 Appendices...