S
PECTROSCOPY
130
copy panel. When selecting a line, clicking the chart once creates a line
from the clicked position to the centre of the image.
Captures the measurement currently displayed in the Spectroscopy window
in a measurement document, and displays it in a separate window.
If
is clicked during the measurement, a copy is generated when the
measurement in progress is finished. During the measurement, the button
remains pressed. The capture process is cancelled by clicking
a second
time.
The measurement documents are labelled automatically with increasing
numbers (i.e. Image1, Image2,...). When leaving the program you will be
asked if you want to save the unsaved measurement documents.
The Spectroscopy panel
Modulation (Easy, Standard, Advanced)
Modulated output (Easy, Standard, Advanced)
The number of available modulated outputs depends on the scan head and
the number of installed modules operating modes that are available. Possi-
ble values are: ‘Z-Axis’, ‘Tip Potential’ and the names of the User Outputs.
Start value/End value (Easy, Standard, Advanced)
The range over which the Modulated output is changed. The ‘Spec For-
ward’ data is measured from the Start to the End value, the ‘Spec backward’
data is measured in the opposite direction. The ‘Spec forward’ data is always
measured before the ‘Spec backward’ data.
Summary of Contents for easyScan 2 AFM
Page 1: ...Operating Instructions easyScan 2 AFM Version 1 6...
Page 7: ...7...
Page 86: ...AFM THEORY 86 Scanner coordinate system x y...
Page 159: ...THE SCRIPT CONFIGURATION DIALOG 159 This page was intentionally left blank...
Page 160: ...AUTOMATING MEASUREMENT TASKS 160 This page was intentionally left blank...
Page 161: ...THE SCRIPT CONFIGURATION DIALOG 161 This page was intentionally left blank...
Page 163: ...163 Window Operating windows Imaging 120 Positioning 115 Report 153 Spectroscopy 128...
Page 164: ......