English
13
Trouble
Cause
Remedy
The motor does not
rotate.
The motor handpiece has an
internal failure.
Contact your Authorized NSK Dealer.
Short-circuit in the ON/OFF
key.
No battery voltage.
<Motor handpiece>
12
Specifications
Rated input
DC 3.6V 0.6A
Rated output
3W
Speed
500 - 2,500rpm
Charging time
Approx. 90 minutes
Dimensions
L150xW25xD25mm
Weight
90g
<Motor handpiece>
Rated input
DC 5V 0.6A
Dimensions
W80xD105xH37mm
Weight
120g
<Stand>
Rated input
100 - 240V AC 0.4 - 0.2A
Dimensions
W46.3xD80xH48.8mm
Weight
155g
<AC adaptor>
Temperature
Humidity
Atmospheric Pressure
Use Environment
10 - 40°C (50 - 104°F)
30 - 75% RH
-
Transportation and Store
Environment
-10 - 50°C (14 - 122°F)
10 - 85% RH
500 - 1,060hPa
*
Use the product in a place where the motor handpiece is free from moisture.
** Using in conditions outside the limits above may cause malfunction.
13
Classification
•
Type of protection against electric shock:
-
Class II equipment
•
Degree of protection against electric shock:
- Type B applied part (Applied part: Disposable handpiece, Motor handpiece)
•
Method of sterilization or disinfection recommended by the manufacturer:
- See "6-1 Cleaning the motor handpiece and stand"
•
Degree of protection against ingress of water as detailed in the current edition of IEC 60529:
- Motor handpiece … IPX0
•
Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide:
- Equipment NOT suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or
nitrous oxide.
•
Mode of operation:
- Intermittent operation
Summary of Contents for NSK iProphy Air
Page 1: ...002 2019 XX XX 001 N ...
Page 35: ...002 2019 XX XX 001 N ...