69
English
CAUTION
・
Be sure to connect the ground wire to the earth ground. Insufficient grounding could cause an
electric shock or fire, and malfunction.
・
When connecting the power cord to the AC Power Input Terminal Block, tighten the screw securely.
Loose Terminal Screws to the AC Power Input Terminal Block will cause overheating leading to
damage and fire in the CONTROLLER.
The power cord includes 3 colors of cords: white, black and yellow/green. Make sure to connect each cord to the
AC Power Input Terminal Block, etc. securely. (Table. 2, Fig. 8)
Table. 2
10 - 1 Wiring of the power cord
Wire Color
AC Power Input Terminal
Block connection position
White
Neutral line (N)
Black
Live line (L)
Green/Yellow
Ground wire
Neutral line
(N)
AC100 - 240V
Live line (L)
Ground wire
Fig. 8
CAUTION
・
Reduce the risk of unintentional starting. Make sure the Main Power Switch
⑨
is in the OFF position
before connecting the CONTROLLER or plugging the system in.
・
When installing the CONTROLLER , provide space of approximately 10cm around the CONTROLLER
for easy access to the air inlet and the Power Cord Connector.
(1) Install the Power Cord Hook
⑫
to the Fixing Power Cord Hook Bar
⑪
(Refer to Fig. 9).
(2) Insert the female plug into the main power inlet box (Refer to Fig. 10).
(3) Secure the Power Cord Connector by use of Power Cord Hook
⑫
(Refer to Fig. 11).
10 - 2 Power Cord Connector Connection
Fig. 9
Fig. 10
Fig. 11
Fixing Power
Cord Hook Bar
⑪
Main Power
Switch
⑨
Power Cord Hook
⑫
Power
Cord
Connector