NB203QR Operation Manual   

CO2 inc Roller

 

 

 

 
 
 
     

Model : NB-203QR 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for NB203QR

Page 1: ...1 NB203QR Operation Manual CO2 incubator Roller Model NB 203QR...

Page 2: ...ON 3 2 TRANSPORTATION STORAGE AND LOCATION OF INSTALLATION 5 3 PREREQUISITE AND CONFIGURATION 7 4 FEATURES AND SPECIFICATIONS OF THE UNIT 9 5 CONTROL PANEL 10 6 OPERATION 11 7 SAFETY SWITCH 16 8 ROLLE...

Page 3: ...A space between the product and the plug must be 30cm at least The power outlet must be only for this product Using various products simultaneously can cause a fire Clean the power plug with a dry to...

Page 4: ...fire 1 4 Precaution for use You must not disassemble fix and remodel the product by yourself You can damage the product throughout a fire and malfunction or get a property loss as well as experimenta...

Page 5: ...nsportation DO NOT try to slide or tilt the unit Lift the unit on each corner of bottom with the aid of 2 people Permissible ambient temperature range for transport 10 to 60 2 2 Storage Do not keep it...

Page 6: ...ase of danger Install the unit at a flat surface free from vibration and in a well ventilated location If the ground is not flat it may cause an excessive vibration of the product When you move the pr...

Page 7: ...heat or cold air to incubator 3 Directly Sunlight Exposed to incubator 4 Uneven ground or table head 5 The place where is being vibrated Cleaning before use Before conducting cell culture It is recomm...

Page 8: ...oor Roller apparatus Roller rack fixing pin Water pan Heated by 6 sides direct heating for humidity Outer door heater prevent condensation CO2 inlet Safety switch Power cord Fuse holder NOTE 1 The fig...

Page 9: ...rature Range Ambient 5 to 60 Accuracy 0 25 at 37 Controller Microprocessor Digital PID Humidity 70 80 at 37 CO2 Range 0 to 20 Accuracy 0 1 at 5 at 37 CO2 Increment 0 1 CO2 Sensor IR CO2 Sensor Roller...

Page 10: ...ay LED window 5 CO Gas display window 6 Pilot Lamp for CO2 valve Pilot lamp ON position means inflow of gas 7 Roller RPM display 8 Roller operating pilot lamp ON running OFF not running 9 Roller power...

Page 11: ...roduct at the desired place and check the level in all directions side by side front to back and ground 6 2 Connect the power Prior to connect the power plug make sure that the POWER S W is off 6 3 Co...

Page 12: ...of current temperature and CO2 in chamber will be displayed after below message passed 6 5 Setting temperature a Connect the power then the LED screen will display the temperature in the chamber b Pr...

Page 13: ...FUNCTION 1 Chamber Temperature 2 Door Heater Temperature 3 Glass Door Heater Temperature 4 CO2 Calibration 5 Heating control 6 CO2 gas supply control 7 To apply a new value a Press and hold CAL SET f...

Page 14: ...ibration Channel 3 is purposed to remove excessive water condensing on glass door caused by high temperature difference between chamber and outside When failed to remove water condensing by calibratin...

Page 15: ...sity reaches to 3 From 3 of CO2 to setting value solenoid valve takes control of CO supply until it gets to setting value When above stage is cleared please press the button to save the new value 6 8...

Page 16: ...RPM TIME Set up RPM setting range is from 2 30rpm and time range is from 00 00 99hours 59minutes RPM setting change can be done by UP DOWN button and possible to change setting value during running W...

Page 17: ...on Press SET button once more then last 2 digits of DISPLAY TIMER will be flickering Set the desired time of minute by UP DOWN button Press SET button to save then t SAVE on display will appear Start...

Page 18: ...function of the CO2 controller Reset the alarm CO2 sensor system defective Contact N BIOTEK service The concentration of CO2 or O2 does not reach the adjusted set value Gas inlet defective Replace th...