Page 1: ...CO2 Incubator Operation Manual Model NB 203XXL...
Page 2: ...s NO CONTENTS PAGE WARRANTY 3 GENERAL INFORMATION ON PRECAUTION 4 TRANSPORTATION STORAGE AND LOCATION OF INSTALLATION 6 PREREQUISITE AND CONFIGURATION 8 FEATURES SPECIFICATIONS CONTROL PANEL 11 OPERAT...
Page 3: ...h have been misused neglected accident or mis installation application 1 The free warranty service will be provided once the unit is proved to be defective by wrong workmanship after NBIOTEK or reliab...
Page 4: ...arious products simultaneously can cause a fire Clean the power plug with a dry towel and connect it properly Foreign substances or unsafe connection can cause a fire Do not bend the power cable hard...
Page 5: ...and a Fire 3 Precaution for use You must not disassemble fix and remodel the product by yourself You can damage the product throughout a fire and malfunction or get a property loss as well as experime...
Page 6: ...ge and Location of Installation 1 Transportation DO NOT try to slide or tilt the unit Permissible ambient temperature range for transport 10 to 60 2 Storage Do not keep it at Place in High Humidity Pe...
Page 7: ...e of danger Install the unit at a flat surface free from vibration and in a well ventilated location If the ground is not flat it can cause an excessive vibration of the product When you move the prod...
Page 8: ...generating heat or cold air to incubator 3 Directly Sunlight Exposed to incubator 4 Uneven ground or table head 5 The place where is being vibrated 6 Too narrow to use lift handle at side of bottom an...
Page 9: ...fan Magnetic door sealing Glass door heater Prevent condensation Glass door latch Divided sealing section 2 Glass doors Stainless steel shelf Perforated to improve uniformity Inner door sealing Airtig...
Page 10: ...necting Gas Tubes Find two gas inlet ports at back of incubator and connect the incubator to the CO2 supply using 6mm PVC tubing with installed HEPA filter There is direction of gas injection flow on...
Page 11: ...reservation between insulation and chamber 2 Specification Items NB 203XXL Temp Range Ambient 5 to 60 Temp Accuracy 0 5 at 37 Temp Increment 0 1 Controller Microprocessor Humidity 70 80 at 37 CO Range...
Page 12: ...ting activation at 3 parts where is controlled by 3 each sensors Glass door means the heater around glass door 4 Temperature Display LED window 5 CO Gas Display 6 Pilot Lamp for CO2 valve Pilot lamp O...
Page 13: ...process 1 Place and install the product Install the product at the desired place and check the level in all directions side by side front to back and ground 2 Connect the power Prior to connect the po...
Page 14: ...ply is up to 4 This will be covered more at page 17 4 Power switch on Once power is on digital readout of current temperature and CO in chamber will be displayed after below message passed 5 Setting t...
Page 15: ...AY FUNCTION 1 Chamber Temperature 2 Door Heater Temperature 3 Glass Door Heater Temperature 4 CO2 Calibration 5 Heating control 6 CO2 gas supply control 7 To apply a new value a Press and hold CAL SET...
Page 16: ...calibration Channel 3 is purposed to remove excessive water condensing on glass door caused by high temperature difference between chamber and outside When failed to remove water condensing by calibr...
Page 17: ...From 4 of CO2 to setting value solenoid valve takes control of CO supply until it gets to setting value When above stage is cleared please press the button to save the new value 8 Alarm In order to ac...
Page 18: ...the heater from overheating when the temperature controller is malfunctioning Set the Safety S W higher than setting point The Safety S W has wide deviation Safety S W is the safety device for prevent...
Page 19: ...gers again there is short circuit contact N BIOTEK service Switch defective Replace the switch Gas CO2 concentration in chamber is too high too low Defective function of the CO2 controller Reset the a...