1
NB-203QS
Operation Manual
CO2 Inc 1 Shaker
Model: NB-203QS
Page 1: ...1 NB 203QS Operation Manual CO2 Incubator 1 Shaker Model NB 203QS...
Page 2: ...ON 3 2 TRANSPORTATION STORAGE AND LOCATION OF INSTALLATION 5 3 PREREQUISIT AND CONFIGURATION 7 4 FEATURES AND SPECIFICATIONS OF THE UNIT 9 5 Control Panel 10 6 OPERATION 11 7 Safety Switch 16 8 Shaker...
Page 3: ...en the product and the plug must be 30cm at least The power outlet must be only for this product Using various products simultaneously can cause a fire Clean the power plug with a dry towel and connec...
Page 4: ...d remodel the product by yourself You can damage the product throughout a fire and malfunction or get a property loss as well as experimental loss Do not use the product for different purpose It can c...
Page 5: ...e unit on each corner of bottom with the aid of 2 people Permissible ambient temperature range for transport 10 to 60 2 2 Storage Do not keep it at place in High Humidity Permissible ambient humidity...
Page 6: ...ase of danger Install the unit at a flat surface free from vibration and in a well ventilated location If the ground is not flat it may cause an excessive vibration of the product When you move the pr...
Page 7: ...eat or cold air to incubator 3 Directly Sunlight Exposed to incubator 4 Uneven ground or table head 5 The place where is being vibrated Cleaning before use Before conducting cell culture it is recomme...
Page 8: ...event CO2 leakage Outer door Shaker apparatus Outer stainless door heater prevent condensation and decontamination CO2 inlet Safety switch Power cord Fuse holder NOTE 1 The figure 1 is to show you the...
Page 9: ...5 to 60 Accuracy 0 25 at 37 Controller Microprocessor Digital PID Humidity 70 80 at 37 CO2 Range 0 to 20 Accuracy 0 1 at 5 37 CO2 Increment 0 1 CO2 Sensor IR CO2 Sensor Shaker RPM 30 to 300 rpm RPM Ac...
Page 10: ...s door means the heater around glass door 4 Temperature display LED window 5 CO Gas display window 6 Pilot Lamp for CO2 valve Pilot lamp ON position means inflow of gas 7 Shaker activation lamp this l...
Page 11: ...wer plug make sure that the POWER S W is off Before Switch ON insert water tray filling with distilled water Make sure connection of gas supply And Open CO2 gas cylinder or supply with the pressure of...
Page 12: ...l stop flickering If you don t press SET key lastly after adjusting the new set up value will not be saved at all Set up Temperature range is Ambient 5 60 Normal Note The initial set point for tempera...
Page 13: ...perature 3 Glass Door Heater Temperature 4 CO Deviation Calibration 5 Heating On Off Range Control Factory Setup Only DO NOT CHANGE 6 CO gas supply speed control Factory Setup Only DO NOT CHANGE 7 To...
Page 14: ...libration Channel 3 is purposed to remove excessive water condensing on glass door caused by high temperature difference between chamber and outside When failed to remove water condensing by calibrati...
Page 15: ...This is pre programmed mode before releasing from manufacturer Push UP and DOWN to set the value NOTE Channel 6 is to set starting point of solenoid control for CO supply Factory pre programmed set po...
Page 16: ...nitial alarm if either or both CO2 and Temperature is still in alarm range This is to remind user of the error which continued and not corrected 7 Safety Switch It is the safety device to prevent the...
Page 17: ...ickers you can set the hour through pressing UP DOWN keys and after setting the hour you can set the minute through the same way with setting the hour d Press the SET key after setting the time hour a...
Page 18: ...will stop when TIME ends and END will be displayed At this stage to press START STOP Button is to resume the TIME function same as before If you need another TIME setup re arrange the setting The rem...
Page 19: ...n of the CO2 controller Reset the alarm CO2 sensor system defective Contact N BIOTEK service The concentration of CO2 or O2 does not reach the adjusted set value Gas inlet defective Replace the gas tu...