PC CONFIGURATION
Scroll down the service list, right click on the “Windows
firewall” service and double click.
Go to the “Startup type” drop-down menu and select
“Disabled”. Then press the “Stop” key to stop the
service currently being executed.
Wait until the service is stopped:
Click on “OK” to close the dialogue window.
From the Windows Services list, search for “Windows
update”, stop the service and disable the automatic
startup.
3.5 PC-DEVICE CONNECTION
The device may be connected to a local area network or directly to the Acquisition Server PC. In both cases, the 3D
panel Ethernet connection must absolutely be point-to-point and is always managed autonomously by the software
and the machine. Therefore, it is not necessary to set any IP address on the INTEL PRO 1000 board, as it will be set
autonomously during the software installation procedure described in the following chapters. On the other hand, it may
be necessary to set a static IP address for the network card dedicated to PC – device communication. If this network
card too is characterised by a direct connection (see the following figure), then it will be necessary to set an IP address
compatible with the default address of the device (192.168.1.160).
Summary of Contents for Hyperion X5 2D
Page 1: ...www my ray com Cod 97071218 Rev 01 01 2017 hyperion X5 2D 3D technical manual EN ...
Page 2: ......
Page 4: ...2 EN hyperion X5 2D 3D TECHNICAL MANUAL CONTENTS ...
Page 37: ...35 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION ...
Page 40: ...38 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION PIN 4 signal X ray ON ...
Page 41: ...39 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION ...
Page 45: ...43 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION ...
Page 47: ...45 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION 5 5 MONOBLOC AREA ...
Page 51: ......
Page 52: ......