pH
/
Temperature
Probe
_________________________________________________________________________________________________
Part No. 78088-00
55226.R004.042016
1 of 3
GENERAL
DESCRIPTION
The
MVP
ICON®
pH
/
Temperature
Probe
is
semi
‐
conductor
designed
for
general
purpose
pH
measurement
of
liquids
or
semi
‐
solids
ranging
from
0
–
14
pH.
Temperature
measurements,
ranging
from
0
–
80
°C,
may
be
taken
simultaneously.
The
probe
features
a
silver
/
silver
chloride
/
potassium
chloride
reference
system
and
a
thermistor
with
automatic
temperature
compensation.
Solid
‐
state
componentry
built
into
the
MVP
ICON
instrument
controls
this
sensing
element.
The
MVP
ICON
pH
/
Temperature
Probe
is
designed
for
use
with
the
MVP
ICON
instrument
only.
Any
other
combination
might
cause
loss
of
performance
or
irreversible
damage
to
both
probe
and
instrument.
Part
No:
78088
‐
00
KIT
COMPONENTS
Each
kit
contains
one
MVP
ICON
pH
/
Temperature
Probe
and
a
soft
bristle
brush
for
cleaning
the
sensor
on
the
tip.
TEST
PROCEDURE
A.
Using
the
MVP
ICON
pH
/
Temperature
Probe
Note:
The
MVP
ICON
pH
/
Temperature
Probe
must
be
calibrated
for
pH
prior
to
first
use.
See
B.
Calibration
for
pH
for
details.
(a)
Remove
and
save
the
protection
cap
or
tube
from
the
probe
tip
prior
to
use.
Note:
Reference
gel
may
be
observed
as
a
viscous
material
on
the
tip
of
the
probe.
Some
gel
seepage
from
a
new
probe
is
normal
and
will
not
affect
the
longevity
or
performance
of
the
probe.
(b)
Prior
to
each
day
of
use,
scrub
the
probe
tip
with
a
soft
bristled
brush
(included)
and
deionized
water
to
remove
possible
residues.
Use
a
mild
detergent
if
required.
(c)
If
the
probe
has
not
been
used
for
2
or
more
days,
clean
as
above,
soak
tip
in
saturated
KCI
(potassium
chloride)
solution
(see
OPERATING
TIPS
)
for
10
‐
15
min,
then
soak
in
pH
7.00
or
pH
4.00
buffer
for
at
least
10
min.
(d)
Ensure
the
probe
is
securely
connected
to
the
MVP
ICON
instrument
by
inserting
the
probe’s
connector
with
the
arrows
facing
toward
the
back
of
the
instrument
and
the
white
circular
sticker
facing
the
front.
Insert
probe
into
the
sample.
Follow
instructions
on
the
MVP
ICON
to
complete
reading.
Both
pH
and
temperature
readings
will
be
displayed
on
the
instrument
screen.
Note:
To
disconnect
the
probe,
squeeze
both
sides
of
the
connector
while
unplugging
from
the
MVP
ICON.
(e)
When
using
the
MVP
ICON
pH
/
Temperature
probe
with
semi
‐
solids,
insert
probe
so
that
both
sensor
and
porous
white
plug
on
opposite
side
of
probe
tip
are
contacting
or
immersed
in
the
sample.
Rotate
left
to
right
several
times
and
tilt
to
ensure
constant
contact
with
the
sample.
(f)
To
ensure
correct
measurement
values,
samples
or
buffers
should
be
mixed
well.
This
may
be
done
using
a
magnetic
stirrer
or
by
gently
stirring
the
probe
in
the
sample
for
at
least
5
sec.
You
may
continue
to
stir
while
the
reading
is
taking
place.
(g)
The
probe
should
be
rinsed
thoroughly
with
deionized
water
and
blotted
dry
between
samples.
B.
Calibration
for
pH
The
MPV
ICON
pH
/
Temperature
Probe
features
a
3
‐
point
calibration
using
pH
7.00,
then
pH
10.00,
and
finally
pH
4.00
buffer
solutions.
Note:
Use
a
borate
‐
based
pH
10.00
buffer
(not
a
carbonate
based
buffer)
to
ensure
a
robust
calibration
slope
for
pH
values
above
7.00.
Note:
In
order
to
perform
accurate
pH
measurements,
the
MVP
ICON
pH
/
Temperature
Probe
must
be
calibrated
daily
prior
to
use,
or
when
using
a
different
MVP
ICON
pH
/
Temperature
Probe.
(a)
Rinse
the
MVP
ICON
pH
/
Temperature
Probe
with
deionized
water.
Soak
probe
in
fresh
pH
7.00
buffer
for
at
least
2
min
prior
to
calibration.
(b)
From
the
MVP
ICON,
enter
pH
mode,
press
Device
,
then
press
Calibration
.
(c)
Insert
the
probe
into
pH
7.00
buffer
to
begin.
Repeat
with
fresh
pH
10.00
and
pH
4.00
buffers
when
prompted
on
the
MVP
ICON
screen.
Ensure
buffers
are
well
mixed.
(d)
Rinse
probe
with
deionized
water
and
blot
dry
between
buffers.
C.
Calibration
for
Temperature
The
MVP
ICON
pH
/
Temperature
Probe
uses
a
factory
‐
calibrated
thermistor.
It
only
needs
to
be
re
‐
calibrated
if
it
has
been
used
to
measure
temperatures
20
°C
above
or
below
room
temperature.
The
probe
features
a
single
‐
point
calibration
that
can
be
done
at
any
known
temperature
within
the
operating
range.
Note:
The
calibration
is
most
accurate
when
using
a
known
temperature
close
to
the
sample
temperature
range.
(a)
From
MVP
ICON,
enter
TEMPERATURE
mode,
press
Device
menu,
then
press
Calibration
.