SDTA-01
5BMAINTENANCE
3.
Use the CALIBRATION1 command to adjust the zero reading and follow the
instructions.
4.
Replace the connector on the back panel.
5.
Connect a voltmeter across the same analog signal.
6.
Use the CALIBRATION1 command to adjust the scale factor of the selected
analog channel.
7.
In response to “ENTER THE APPLIED VALUE:” on the MXP2 188, enter the
voltmeter reading and follow the instructions.
NOTE
: for the universal module, it is not necessary to short the input in order to
adjust to zero since the module does that automatically.
5.2 Troubleshooting
This section lists some helpful hints on how to analyze problems that may occur,
because all troubles are unpredictable. When the problem is still not identified after
reading this section, please refer to the Customer service centre.
NOTE: before sending back any defective unit, please make arrangements with your
nearest MULTITEL Customer service centre.