Running the Example HCF Test
90
Landmark™ Test System Operation
Adjust Limits for the Test
Setting force limits just outside the expected values for the HCF test helps prevent specimen and
equipment damage if something unexpected happens during a test run.
1. You can retain the wide displacement limit range set earlier because this example tension test
uses no fixturing that requires protection.
A. From the Station Manager main window, select
Display > Station Setup
, and click
the
Channel Input Signals
button.
B. Select the
Displacement
channel.
C. Select the
Limits
tab.
D. Use the following parameters:
l
Landmark 1 Displacement (Upper Limit, Upper Action)
— 75 mm,
Interlock
l
Landmark 1 Displacement (Lower Limit, Lower Action)
— (- 75) mm,
Interlock