Running the Example HCF Test
Landmark™ Test System Operation
81
4. Set up a force waveform using the Function Generator.
This introduces a force waveform to the specimen.
A. In the Station Manager main window, click the Function Generator button.
B. Set up a force waveform using the following parameters:
l
Channel
— Landmark 1 (or other depending on your configuration)
l
Control Mode
— Force
l
Command Type
— Cyclic
l
Target Setpoint
— +500 N
l
Amplitude
— 250 N
l
Frequency
— 1.00 Hz
l
Wave Shape
— Ramp
l
Compensator
— None
Important:
It is important to have
Compensator
set to
None
. If it is
left on, it can negate the function generator signal.
5. Open a Station Manager Scope.
Click the Scope button in the Station Manager main window.