Running the Example HCF Test
76
Landmark™ Test System Operation
Position the Crosshead
1. Adjust the position of the crosshead using the hydraulic controls.
Item Description
1
Crosshead Lock
2
Crosshead Lift
A. Unlock the crosshead clamps, and then move the crosshead up or down so that there
is about 25 mm (1 in) clearance between the upper grip and the specimen.
B. Once properly positioned, reclamp with the crosshead lock control.
Move the Specimen into the Upper Grip Using Displacement Control
Displacement control is used to move the specimen into the upper grip. However, care must be
exercised to ensure that minimal force is imparted on the specimen during the process as this could
affect test results.