Running the Example HCF Test
Landmark™ Test System Operation
65
D. Use the following parameters:
Important:
The following numbers were appropriate for the example test on a
typical system. You will need to adapt these entries to your specimen, test, and
system.
l
Landmark 1 Displacement (Upper Limit, Upper Action)
— 100 mm,
Interlock
l
Landmark 1 Displacement (Lower Limit, Lower Action)
— (- 100) mm,
Interlock
2. Set the Force Upper (Tension) Limit.
Note:
For this test, MTS has determined that a setting of +5000 N will prevent damage
to the system and the specimen.
Item Description
1
Type here and press Enter
Note:
You must press
Enter
after typing numerical entries for them to be accepted.
A. Select
Station Manager >Display > Station Setup > Channels > Landmark 1
(or other depending on your configuration) > Force > Limits
, click the Channel
Input Signals button, and then click the
Limits
tab.
B. Set the
Upper Limit
to
+5000 N
(a positive number indicates tension). Be sure to
press
Enter
after typing in the limit.
C. Then, in the
Action
drop-down list, select
Interlock
.