Running the Example HCF Test
Landmark™ Test System Operation
111
l
The actuator may have retracted below the lower displacement limit during specimen
installation.
l
The actuator may have retracted when the specimen broke at the conclusion of the test.
l
Your test limits may have been insufficient for the test.
Rather than cover all possibilities, this section addresses the first case only. Adapt this procedure as
necessary to recover from other situations.
1. Check current displacement limits.
A. In the Station Manager application main window, select
Display > Station Setup >
Displacement > Channel Input Signals
.
B. Select the
Limits
tab.
C. Check the current upper and lower limit settings to identify the acceptable
displacement range for the actuator. Adjust the limits if necessary.
2. Set Manual Command to Displacement.
A. In the Station Manager application main window, click the
Manual Control
button.
B. Ensure that Exclusive Control is checked.
C. In the Manual Controls window, select a Control Mode of
Displacement
.
D. Ensure that Enable Manual Command is checked.