Running the Example HCF Test
110
Landmark™ Test System Operation
7. Reset the Lower (Compression) Limit.
A. Navigate to
Station Manager >Display > Station Setup > Channels >
Landmark 1 (or other depending on your configuration) > Force > Limits
.
B. Enter
-50 N
in the
Lower Limit
text box and ensure that
Interlock
is selected for the
Action
.
8. Release the upper grip.
Using the controls on the load frame, release the upper grip.
9. If necessary, repeat the upper grip clamping process.
If Necessary, Recover from a Tripped Displacement Limit