Running the Example HCF Test
Landmark™ Test System Operation
101
Remove Displacement Limits
1. Remove displacement limits.
Because the actuator may retract at the conclusion of the test, begin by removing the
displacement limits.
A. From the Station Manager application main window, select
Display > Station
Setup
, and click the
Channel Input Signals
button.
B. Select the
Displacement
channel.
C. Select the
Limits
tab.
D. Set the
Displacement Upper Action
and
Lower Action
to
Disabled
.
Switch to Running Max/Min Displacement and Force Meters
1. Switch to a Running Max/Min Displacement meter.
Because you are no longer imparting a waveform on the specimen, change the displacement
Peak/Valley meters to Running Max/Min Meters to monitor history during specimen removal.