Page 33
MTI Part Number 7001-0040
Revision 3.1 August 7, 2014
TOOLS Menu – Calibration Wizard Continued
Channel B High range AC calibration
Next, click in the
Voltage Reading
window and enter the
reading of the
voltmeter. Verify that
the correct units (Pk,
Pk-Pk, or RMS) are
selected for the type of
voltmeter you are
using.
This screen is repeated
for the three mid range
voltages (1.4 v, 3.5 v,
and 5.6 v) used.
Summary of Contents for 1510A