M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
MSI-4300 Porta-Weigh Plus/5 Key
•
User Guide
Page 15
CALIBRATE SETUP MENU
Top Menu
Selections
FILTER
LOW, MED, HIGH
UNITS
kg, lb, Ton, Metric Ton
STD CAL
Starts Cal procedure
FINE CAL
For minor calibration adjustments
CAL RCAL
For emergency calibration only
STANDARD
INDUSTRY, NIST, OIML, ALASKA, CANADA, UNITED
KINGDOM, LEGAL-FOR-TRADE, AUSTRALIA
AUTO OFF
ENABLED
, DISABLED
AZM
ENABLED
, DISABLE
REMOTE
ENABLED, DISABLED
LIGHTING
AUTOMATIC
, ON, OFF
CONTRAST
(eight steps with numerics at “123456” and message reading
“CONTRAST”)
EXIT
Push ENTER (TEST) to quit setup
HI SET PT
HI VALUE
DISABLED, GROSS, NETGROSS
RESPONSE
NORMAL, BLANKLCD, BLINKLCD
EXIT
LO SET PT
LO VALUE
INPUT WT
MODE
DISABLED, GROSS, NETGROSS
RESPONSE
NORMAL, BLANKLCD, BLINKLCD
EXIT
TOTALIZE
MANUAL, AUTO
STANDARD CALIBRATION
Standard Calibration is to be used for normal test weight or cal fixture calibra-
tion of the Porta-Weigh Plus.
1) Enable Calibration by pushing the key behind the seal port.
2) Use the SELECT (NET/GROSS) key to scroll to the “STD CAL”
message. Push ENTER (TEST).
3) The message reads “Ø SCALE”. Remove all weight from the scale.
When motion ceases push ZERO. Assuming the detected zero weight is