background image

2

Ⅲ. Components and functions

1

2

3

4

5

1. LED module

containing fluorescence filter cubes, filters, lever used to exchange colors.

2. Light source

LED light source has different wavelength LED lamps.

3. Fixed sheet metal: used to connect and fix LED module to the microscope.

4. Lenses tube

connecting LED module with light source and for lighting.

5.

Power control box

exchange current and voltage,connecting power,control light ON/OFF and

brightness.

Ⅳ. Installation

1. Loosen the fixing screws of waterproof baffle on the left and right sides of Nikon TS100 microscope

host (as shown in Figure 01) with an internal hexagonal screw driver, and remove the plastic baffle plate

to obtain the fluorescent module installation area.

Figure 01

Summary of Contents for MI-BGU-LED-TS100

Page 1: ...MI BGU LED TS100 LED Fluorescence Illuminator User Manual ...

Page 2: ...MI BGU LED TS100 LED fluorescence illuminator ...

Page 3: ...er to ensure your safety prolong the life of this unit and maintain it properly please read this manual carefully before operating this unit Warning Do not use or place the instrument in the place with high temperature humidity or dust for a long time Suitable working temperature 5 to 35 Suitable relative humidity 20 to 80 25 Note do not immerse the instrument in water or solvent Note do not place...

Page 4: ...Content Ⅰ Introduction 1 Ⅱ Main specification 2 Ⅲ Components and functions 2 Ⅳ Installation 2 4 Ⅴ Operation 4 5 ...

Page 5: ...rvation Ⅱ Main specification 1 Specification sheet Excitation group LED wavelength Excitation wavelength Dichroic mirror Emmision filter B 470 475nm 450 490nm 505nm 515nmLP G 530 535nm 510 550nm 565nm 575nmLP UV 365nm 330 380nm 400nm 420nmLP 2 Optional specification Excitation group LED wavelength Excitation wavelength Dichroic mirror Emmision filter UV BP 365nm 340 390nm 400nm 450 65nm B BP 470 4...

Page 6: ...roscope 4 Lenses tube connecting LED module with light source and for lighting 5 Power control box exchange current and voltage connecting power control light ON OFF and brightness Ⅳ Installation 1 Loosen the fixing screws of waterproof baffle on the left and right sides of Nikon TS100 microscope host as shown in Figure 01 with an internal hexagonal screw driver and remove the plastic baffle plate...

Page 7: ...nstallation area of the fixed metal plate from the side as shown in Fig 03 There is a screw hole bit on the edge of both ends of the fixed metal plate Use the fixed screw to connect the metal plate and the module body and lock and fix it as shown in Fig 04 Figure 03 Figure 04 4 Insert in the fixing lenses tube from the bottom rear end of the microscope host connect the main body of the fluorescent...

Page 8: ...he two hexagon socket fixing screws on the lens fixing cylinder as shown in Fig 05 Figure 06 7 Connect the light source control box with the light source illuminator of the module and the installation is completed as shown in Figure 07 圖 07 Figure 07 Ⅴ Installation 1 Turn on the power and turn on the power control box switch Figure 08 Figure 08 ...

Page 9: ...ntions 1 The brightness of the LED in the illuminator can be adjusted freely according to the difficulty of excitation of the observed sample generally lower than the full load current is beneficial to prolong the service life of the LED lamp bead if allowed it is recommended to adjust the brightness below full load for use 2 During fluorescent observation the eyepiece should not directly face the...

Page 10: ...to the smoothness of the module position otherwise the uniformity of the light may become worse This power supply must strictly use the special adapter of 12V 2A and the customer who uses other adapters shall be responsible for the equipment damage ...