Entering System Parameters
GB
SUPREMATouch
Status Texts
Fig. 31 Status texts
Use this window to define sensor type specific texts for specific signal ranges.
These are displayed in the measuring value list with a letter
F
prefix. (e.g.
F:OpticErr
). Texts can
be defined for all remote sensors provided that ranges are specified for them. Texts can be
defined arbitrarily, the same texts are allowed for more than one sensor.
In addition, the user can freely define the signal ranges in the range from 0–400 mV for change-
able remote sensors. However, the signal ranges must not overlap.
UA
idle
(Limit for
signal fail)
Number
(integers)
Adjusting
range: 50–
350, must
be < UA
min
This field displays the maximum signal UA for the status
signal fail
. Measuring values below this limit value are
displayed as
signal fail.
This value can only be set for the three user-changeable data
sets.
UA
over
(Limit for
“overflow”)
Number
(integers)
Adjusting
range:
2000–2200
This value defines the sensor signal UA of the remote sensor
for the measuring range overflow display. Measuring values
above this value are displayed as
overflow
.
This value can only be set for the three user-changeable data
sets.
T
supp
(Warm-up
period)
Number
(integers)
Adjusting
range: 10–
300
The warm-up time in seconds can be set here. This specifies
how long a sensor will remain in the warm-up status (display
suppressed
) after it has been switched on. This time is neces-
sary because different sensors need a different length of time
to warm up and the correct measuring value is displayed.
This value can only be set for the three user-changeable data
sets.
OK
Button
By tapping this button, the completed settings are applied to
the selected head.
Cancel
Button
By tapping this button, the completed settings of the selected
head are canceled.
Field
Field Type Function
Summary of Contents for SUPREMATouch MCP 20
Page 1: ...MSAsafety com Operating Manual SUPREMATouch Fire and Gas Warning Unit Order No 10121863 05...
Page 8: ...MSAsafety com User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Page 72: ...MSAsafety com Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Page 232: ...232 Dimensions GB SUPREMATouch 18 Dimensions 18 1 Rack...
Page 233: ...233 Dimensions GB SUPREMATouch 18 2 Rail mounted Modules MRO8 TS Module MRO16 TS Module...
Page 235: ...235 Dimensions GB SUPREMATouch MRO20 8 TS Module 1 3 2 69 90...
Page 236: ...236 Dimensions GB SUPREMATouch MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Page 237: ...237 Dimensions GB SUPREMATouch MRC TS Module MGT40 TS Module...
Page 238: ...238 Dimensions GB SUPREMATouch MHD TS Module MAT TS Module...
Page 241: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose...