38
Entering System Parameters
GB
SUPREMATouch
Relay Outputs Window
Parameter values for relay outputs can be viewed and changed here.
The functions of window are similar to the
Relay Output
assignment window described in the
previous section. There, starting from a particular input, a connection to a relay output was made.
In this menu, the setting conditions are configured starting from a particular relay output.
The behaviour of a relay output depends on its parameter settings and the settings of the appro-
priate measuring points.
Fig. 20 Relay output
The Relay Outputs window is divided into three subwindows:
• Information
• Sensor connections
• Relay connections
The following paragraphs describe the functions of the individual window fields and buttons. The
first fields described are identical in all three subwindows.
All parameter changes using the menu elements described below apply to the relay
output selected in the
Relay
field.
Field
Field Type
Function
Relay
Selection
Contains a list of available relay outputs. As the first 8 relay
outputs of the system are tied to the common messages,
the first relay output which can be configured is No. 9.
After an output number is selected, the rest of the window
is filled, if settings have already been entered for that
output. This input field can be accessed without a key
switch or password if an output is selected for which param-
eters have already been set. If an output is selected that
has not been used before, authorisation with password or
key switch is required. If an output is displayed that has not
yet been configured, default values are used as the prelim-
inary settings for the certain fields. This makes it easy to
copy the settings from one output to another.
Summary of Contents for SUPREMATouch MCP 20
Page 1: ...MSAsafety com Operating Manual SUPREMATouch Fire and Gas Warning Unit Order No 10121863 05...
Page 8: ...MSAsafety com User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Page 72: ...MSAsafety com Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Page 232: ...232 Dimensions GB SUPREMATouch 18 Dimensions 18 1 Rack...
Page 233: ...233 Dimensions GB SUPREMATouch 18 2 Rail mounted Modules MRO8 TS Module MRO16 TS Module...
Page 235: ...235 Dimensions GB SUPREMATouch MRO20 8 TS Module 1 3 2 69 90...
Page 236: ...236 Dimensions GB SUPREMATouch MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Page 237: ...237 Dimensions GB SUPREMATouch MRC TS Module MGT40 TS Module...
Page 238: ...238 Dimensions GB SUPREMATouch MHD TS Module MAT TS Module...
Page 241: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose...