SingleLine SCOUT
12
Cleaning and Maintenance
GB
5
Cleaning and Maintenance
5.1
Cleaning
WARNING!
Do not use any organic cleaning material for the cleaning of the individual components, such as nitro
thinners, alcohol, white spirit, petrol etc.
The device components should be cleaned after each operation and, if badly contaminated, with a
damp cloth (tepid water) and a little detergent. The device does not need to be removed from the pneu-
matics for cleaning.
5.2
Maintenance
The device is largely maintenance-free. It is recommended that the device is subject to regular visual
inspections to check whether there are any faults or damage.
If there are any faults, the device must be replaced. The device must not be repaired by the user.
Repairs must only be carried out by the manufacturer or by a repair service authorized by the manu-
facturer.
Exchanging the Batteries
Due to storage and delivery times, it might be necessary to exchange batteries before putting the
device into first service.
The charging status of the batteries is monitored by the device. Low battery condition is notified. The
batteries must be changed when notified to do so. If the batteries are empty, the device does not start.
The red battery life indication blinks every 30 seconds.
It is recommended that the batteries be changed at least once a year or if indicated to do so by the
LED or on the display.
The device is designed to provide an operational time of 500 hours or 1 year stand by, the operational
time might vary depending to incident conditions and usage.
WARNING!
Remove batteries if device is stored for more than one year.
Never change batteries in hazardous areas. There is a danger of explosion since the batteries can
cause sparks when being changed! Danger of injury!
Check the polarity of the batteries when replacing. Incorrect polarity connection can lead to damage to
the device or impairment of the operability of the device.
NOTICE
The maximum locking torque for the battery cover is 0.7 Nm. Do not overtighten, otherwise the inserts
can be pulled out!
The device shall only be operated with the following battery types:
•
Energizer L91 Ultimate Lithium
•
VARTA 6106 (Professional) Lithium
(1)
Open the battery compartment with a Torx T10 spanner.
(2)
Replace the batteries with new ones.