Connection of the external control via 4-20 mA input signals
The signal inputs built a 4-bit binary number:
I4 – I3 – I2 – I1: 0-11 mA=open=0;
11/12-20 mA=closed=1.
Status of
external signal
source
Status number
Description
I4 I3 I2
I1
-
-
0 0 0
0
0
Automatic sampling point switching
0
0 0
1
1
Analyzer is sampling the point SP1 (*1, *2)
0
0 1
0
2
Analyzer is sampling the point SP2 (*1, *2)
0
0
1
1
3
Analyzer is sampling the point SP3 (*1, *2)
0
1 0
0
4
Analyzer is sampling the point SP3 (*1, *2)
0
1
0
1
5
Analyzer is sampling the point SP4 (*1, *2)
0
1
1
0
6
Analyzer is sampling the point SP5 (*1, *2)
0
1
1
1
7
Analyzer is sampling the point SP6 (*1, *2)
1 0 0
0
8
Analyzer is sampling the point SP7 (*1, *2)
1
0 0
1
9
Analyzer is sampling the point SP8 (*1, *2)
1
0 1
0
10
Analyzer is sampling the point SP9 (*1, *2)
1
0
1
1
11
Analyzer is sampling the point SP10 (*1, *2)
1
1 0
0
12
Analyzer is “stand-by” (*3)
1
1
0
1
13
Analyzer is “stand-by” (*3)
1
1
1
0
14
Analyzer is “stand-by” (*3)
1
1
1
1
15
Analyzer is “stand-by” (*3)
-
-
-
-
I4
+
I3
I2
I1
+
+
+
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...