10.3.
Analyser states
The IO module, of the analyser, can read out the status of the analyser. To set this
feature, open the path:
E
XTRAS
/
A
NALOG OUTPUT DETAILS
(see pictures below).
Each 1 mA steps represent a function. The chart below shows the different states:
#
mA-Signal
Function
0
4,0 mA
Power-On / self test
1
5,0 mA
Measuring sample point 1
2
6,0 mA
Measuring sample point 2
3
7,0 mA
Measuring sample point 3
4
8,0 mA
Measuring sample point 4
5
9,0 mA
Measuring sample point 5
6
10,0 mA
Measuring sample point 6
7
11,0 mA
Measuring sample point 7
8
12,0 mA
Measuring sample point 8
9
13,0 mA
Measuring sample point 9
10
14,0 mA
Measuring sample point 10
11
15,0 mA
Zeroing
12
16,0 mA
Gas sampling (from any sample point)
13
17,0 mA
System Alarm (measurement halted)
14
18,0 mA
Manual calibration in progress
15
19,0 mA
Automatic calibration in progress
16
20,0 mA
Stand-By
10.4.
Configuration of the external control (Option: IO module)
This feature requires an I/O module (optional) and the function must be activated
.
This feature can be used for the external control of the analyser. With the help of
the external control follow operations can be done:
•
Externally controlled sampling point selection,
•
Stand-by.
The commands will be given by a 4-bit binary number, which will be built through
four external signals. The pins for the signal are shown in the sketch below. It exists
two different types to set the four pins:
•
Potential free relay contacts.
•
4-20 mA signal inputs.
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...