Dilution-unit
The dilution-unit, inside the analyser dilute the sample gas flow with a define ratio.
The unit is preadjusted. The operator must do a fine calibration of the device, to
improve the accuracy of the measurement. For the fine calibration it is
indispensable to tubing the vent and all sample gas inlets, before making the actual
calibration. The correct order for the fine calibration is shown in the flow chart.
Furthermore, it can be necessary due to abrasion and contamination effects to
recalibrate the dilution-unit.
9.1.
Correct installation of the SWG100
See chapter 7.5.
9.2.
Tubing of all sample gas inlets
See chapter 7.6.
9.3.
Installation of the calibration gas cylinder
For the calibration of the dilution-unit, MRU recommends a gas cylinder with a min.
concentration of 500 ppm, from the target gas.
Correct installation of the SWG-100
Tubing of the vent
Tubing of all sample gas inlets
Installation of the calibration gas cylinder
Fine calibration of the dilution-unit
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...