Power up and commissioning the analyser
After installation of the analyzer few steps should be processed in order to operate the
instrument properly.
The graphic below shows the few steps, which should be done, to power up the analyser
correctly.
START
Supervise the atmosphere with an external certified
combustion gas detector
Open the case door of the analyser
Power up the circuit breakers
Press the power switch from the remote power supply box
The analyser will start
Make the analyser commissioning on the operation unit.
-
Language
-
Date and time of the instrument
-
Configure sample points
-
Configure analog in and outputs of the I/O modules
-
Configure the alarm relays
-
Configure the Modbus RTU.
CLOSE and LOCK the cabinet door of analyser
END OF PROCESS